X-Rite eXact™ Instrument
98
Technical Specifications
SPECTRAL ENGINE
Spectral analyzer: DRS spectral engine
Spectral range: 400 nm - 700 nm with 10mm interval
OPTICS
Measurement geometry: 45°:0° ring illumination optics, ISO 5-4:2009(E)
Measurement aperture: 1.5 mm, 2 mm, 4 mm or 6 mm
Light source: Gas filled tungsten (illuminant type A) and UV LED
REFLECTANCE MEASUREMENT
Measurement Conditions: According to ISO 13655:2009):
• M0: Unpolarized, No filter, UV included
• M1: D50
• M2: UV excluded Filter
• M3: Polarization Filter
Calibration: Automatic on integrated white reference
Inter instrument agreement: Average:0.25 ΔE*ab
, Max: 0.45 ΔE*ab (M3 0.55 ΔE*ab)
(Measurements using X-Rite manufacturing standards at a
temperature of 23°C +/- 1°C, 40-60% RH for all
measurement modes on 12 BCRA color tiles and a white
ceramic reference (D50, 2°))
Short term repeatability
White: 0.05 ΔE*ab
, (standard deviation) White BCRA
(Error compared to mean value of 10 measurements
every 5 seconds)
Short term repeatability
Density: +/-0.01 D
for CMYK measurements
(Status E or Status T measurements)
(Maximum error compared to mean value of 10
measurements every 5 seconds at 2.0 D, except for M3
Yellow at 1.7 D)
OPERATING CONDITIONS
Operating Temperature Range: 50° to 95°F (10° to 35°C)
30-85% relative humidity maximum (non-condensing)
Storage Temperature Range: -4° to 122°F (-20° to 50°C)
Comments to this Manuals