X-Rite eXact™ Instrument
27
Basic Measurement Procedure
1. Clear the sample surface of any dirt, dust, or moisture.
2. Select the tool and function.
3. Position the target window over the sample to measure. If
possible, place the entire instrument on the sample.
4. Press the instrument firmly to the target base. Measuring
appears in the display along with the Measurement
Condition selected.
5. Hold steady until “Complete!” and measurement data is
displayed. This is an indication that the measurement was
successful.
6. Release the instrument and view measurement results.
An unsuccessful measurement will be indicated by an error
message. See the Troubleshooting section for more
details.
Calibration
IMPORTANT: The calibration plaque is dramatically affected by smudge marks and dust, and
must be kept clean. It is recommended that the instrument is cleaned regularly. Refer to the
Cleaning section later in this manual for procedure on cleaning the calibration plaque and optics.
The instrument has an integrated calibration plaque and will calibrate automatically when
required. When calibration is required before a measurement, the instrument will prompt you.
This occurs when the instrument is in the open position (not locked) and the previous calibration
has timed out. This can also occur when the measurement condition switch is changed and a
calibration has not been performed in that position or has timed out.
The calibration plaque is located directly below the optics when the instrument is in the open
position (not locked). When a measurement is taken, the calibration plaque swings back out of
the way of the optics.